| Product Code | NE4-SB |
| Max Tank Capacity | 60 Litres |
| Temperature Range | Ambient + 5°C - 99°C
Accessory DC1-300 provides cooling from ambient to -20°C |
| Temperature Control | PID Digital, LED display showing actual or set temperature, resolution 0.1°C |
| Temperature Alarm | Under/over temperature alarm is automatically set 4˚C above and 4˚C below set point |
| Sensitivity | ±0.05°C |
| Timer | Visual indicator and audible end message - variable 00:01 to 99:59 hh:mm |
| Heater Power (220-240V) | 2000W |
| Voltage | 230V / 50-60Hz |
| Dimensions | |
| External (mm) | 122w x 240d x 335h |
| Tank Mounting (mm) | Double clamp fitting accommodates tank max wall thickness 40mm |
| Tank Water Levels (mm) | Example : with overall tank depth 150mm / max water level 130mm / min 80mm |
Features and Benefits
Safety Features
• The under/over temperature alarm is automatically set 4˚C above and 4˚C below set temperature. When in alarm condition, the indicator illuminates and actual bath temperature is shown. If bath is over temperature, both heating and stirring are switched off. Once liquid temperature has fallen or risen above/below the alarm setting, the indicator clears and actual bath temperature is shown.
• Features a low liquid level float switch, which ensures heating is switched off if the liquid falls below the recommended level – display will show “FiLL”. Heating is switched off but stirring continues. To reset, carefully top up the liquid to the recommended level.
Energy Consumption
0.5kWh @ 37°C
The lower the figure, the less energy used. Information supplied for size and range comparison, guidance only and measured consumption at set point. Please remember that the energy consumption may vary due to factors such as location, ventilation, ambient temperatures and loading.
Controls
Featuring a bold LED display of measured temperatures in real-time and in increments of ±0.1°C, using a precision PID temperature controller. Setting for water temperature only or combination of temperature and time. Timer setting variable in 1-minute increments from 00:01 (hh:mm) to 99:59 (hh:mm) with “end” message on completion of the timed period. Time and temperature settings are retained in the memory, even after the unit has been turned off.
Heating/Cooling Information
For cooling a water bath use an accessory immersion cooler DC1-300 refer to Accessories. Where bath temperatures are below ambient we recommend a mix of glycol and water. We also recommend a lid or layer of spheres for insulation. The water bath should be used in ambient temperature 20°C. For water temperatures above 60°C we recommend that an accessory lid or layer of spheres are used for optimum performance. This will reduce evaporation and a lid will keep the samples and water free from contamination. Optimum performance saves energy by reducing electricity consumption and running costs.