Instruction Manual
NE1 Series Unstirred Analogue Baths
The Clifton RangeĀ® is part of Nickel-Electro Ltd a family firm based in Weston-Super-Mare which was incorporated as a limited company in 1941 but its roots can be traced back to 1935 when the business first started. Now in its 3rd generation of family members, the company prides itself on being a strongly established, independant British manufacturer.
Thank you for purchasing this piece of Clifton RangeĀ® temperature control equipment. To get the best performance from your equipment and for your own safety please read these instructions carefully before use.
General Notes
1. This product is designed for laboratory use only. Always follow good laboratory practice.
2. Fill the tank prior to connection to electrical supply.
3. Use caution when topping up/draining the tank. If this product is not used in accordance with these instructions then basic safety protection may be affected.
4. The mains supply cord fitted to this product is heat resistant and should be replaced with an equivalent type by a qualified electrician.
5. Ensure that the power supply has a safety earth (ground) terminal.
6. Ensure that the mains switch and power supply connector are accessible during use.
7. Before using any cleaning or decontamination method please refer to the Maintenance and Cleaning section to ensure the proposed method will not damage the unit.
8. Connect only to a power supply with the corresponding voltage to that specified on the rating label positioned on the rear of the unit.
9. Do not block ventilation slots during use and always follow installation instructions.
10. Ensure substances being heated present no risk of a hazard (explosion, implosion or release of toxic or flammable gases) or that these have been addressed. When heating substances where liberation of gases occurs suitable extraction should be used.
11. Use only liquids specified within this Instruction Manual within their specified temperature range.
12. Drain before moving the bath.
13. We recommend using a lid above 60ĖC. Take care when lifting the lid as steam and hot vapours can cause scolding.
14. Use a thermometer to check the temperature – do not touch the liquid.
Location
The product must be placed on a smooth, level and sturdy work surface, preferably near a drain for emptying. Use in a ventilated room. Suitable for use in ambient temperatures 5°C to 40°C with a maximum humidity 80% (temperature 31°C) decreasing to 50% (temperature 40°C).
DO NOT block or restrict ventilation slots.
DO NOT place directly next to hot heat surfaces. ENSURE that there is sufficient space around the product to allow it to provide optimum temperature control.
Unpacking
Remove the product from its packaging and retain over the warranty period. Contents consist of:
⢠Bath.
⢠Stainless steel false base.
⢠Power lead.
⢠Instruction manual.
Place the false base into the bath, with the legs facing down. Fit the power lead into the socket at the rear.
Safety
Power Lead and Connection To Electrical Supply
Check the electrical supply is compatible with the rating label. IF IN DOUBT CONSULT AN ELECTRICIAN. THE PRODUCT MUST BE EARTHED!
Where the mains supply or plug connection differs refer to local regulations or consult an electrician.
Liquid Level and Suitable Liquids
Maximum liquid level – 130mm/shelf fitted working depth 115mm
Suitable Liquids
Operating temperatures from ambient +5ĖC to 99ĖC, for general use we recommend:
⢠Distilled Water.
⢠Heat Transfer Liquid (the LB range is formulated for temperatures from -45ĖC to 90ĖC and provides complete protection from freezing and algae growth and safeguards against corrosion).
⢠Virkon dissolved in distilled water has proven efficacy against bacteria (inclmycrobacteria), viruses, spores and fungi in a variety of independent tests using different protocols.
⢠Bath Armor – high thermal conductivity allows this material to replace heating/ cooling fluids in water baths (see accessories for full range).
Operating Instructions
NE1 Series Unstirred Thermostatic and Shallow Baths
- Ensure the stainless steel false base is fitted and fill the bath with water following Liquid Level Instructions. SWITCHING THE BATH ON WHEN DRY WILL DAMAGE THE HEATER AND MAY INVALIDATE YOUR WARRANTY.
- Connect the mains power lead to an electrical supply.
- Turn the bath on using the green mains power switch located on the front.
- Adjust the blue temperature dial to the desired setting. DO NOT OVERTURN THE DIAL PAST ITS STOP POINT. DOING SO MAY INVALIDATE YOUR WARRANTY.
- The lockable dial can be locked into position to prevent accidental adjustment. Using the allen key supplied (M3) tighten the fixing screw. Dial is now locked into set position.
- Please note; screw can be loosened off or removed completely if you wish to adjust the dial freely.
- Amber heater indicator will illuminate to show heater activity.
- Allow the water temperature to stabilise in the bath, and take an actual temperature reading using a thermometer. Adjust the temperature controller accordingly.
Care and Maintenance
DISCONNECT THE BATH FROM THE POWER SUPPLY PRIOR TO CLEANING.
Basic Cleaning
The Clifton RangeĀ® bath should be emptied at the end of each day, then for interior surfaces:
Wash ⢠Rinse ⢠Dry
Scale Build Up: In hard water areas limescale can build up and reduce the efficiency of the water bath. Cleaning at the end of each day can prevent this but periodically it may be necessary to descale the bath. Add 1 litre of vinegar to the normal capacity of water and heat for 1 hour to 50ĖC.
Occasionally small spots may occur in the bath which resemble rust. These are usually small ferrous particles which have oxidised. Using a stainless steel cleanser such as Bar Keepers FriendĀ® will remove these marks aswell as mineral deposits.
Exterior Anti Bacterial Painted Surfaces
The āanti-bacterialā paint finish inhibits the growth of bacteria. It has been tested by independant specialist houses using internationally recognised test methods and proven to be effective against a wide range if bacteria including Escherichia Coli and Staphylococcus Aureus (MRSA).
We recognise hygenic coatings are part of a controlled approach to a cleaner working environment. Within the paint formulation is an active ingredient with proven anti- bacterial properties which is maintained throughout its life span. In a laboratory environment this is one less source of contamination. Unlike detergents the anti-bacterial paint finish does not offer an instantaneous action, but is intended for long term general protection against bacterial growth.
Moisture on the painted surface is neccessary for the bacterium to absorb the agent and be affected by it. The coating is therefore less active in very dry conditions although moisture in the atmosphere will maintain some activity. Areas where moisture is trapped are difficult to clean and allow bacteria to proliferate but these areas are most active for the anti-bacterial coating improving defence against bacterial growth.
Decontamination of Equipment
We recommend VIRKON tablets for the safe and rapid disinfection of equipment. Please follow the User Instructions carefully. Virkon solution only requires 10 minutes contact time to be effective. Care should be taken with stainless steel tanks and it is important that virkon solution is not left in contact with metal surfaces āfor longer than is neccessaryā.
We recommend PERASAFE powder for the safe and rapid chemical sterilant of equipment. Please follow the User Instructions carefully.
Please contact your distributor or Day-Impex Ltd for further information relating to these products.
Warranty Terms and Conditions
- Nickel Electro Ltd warrants to the Customer that the product purchased is free from defects in materials and workmanship.
- Provided the terms of payment are duly complied with, Nickel Electro Ltd undertakes to remedy any original defects arising from faulty materials or workmanship, in any goods manufactured/supplied by Nickel Electro Ltd, which under proper and normal conditions of use, may develop within a period of three years from the date of delivery.
- In the case of components which by their nature of application have an unpredictable life, this guarantee shall only be to the extend of the guarantee given by the manufacturers of these articles.
- Nickel Electro Ltd will accept no liability, where in the opinion of the company the defect has been caused by damage due to the Customers failure to follow operating instructions, correct installation, wear and tear, or damage due to the use of spare parts other than those spare parts of Nickel Electro Ltd or which are recommended by Nickel Electro Ltd, the defect has been caused by alterations or repairs being undertaken by a person(s) other than an authorised representative of Nickel Electro Ltd.
- Any damage claim must be in writing, and give the serial number and description of the goods, order number and date of delivery, and will not apply where any names or serial numbers or other information which may be attached to or inscribed upon the goods have been removed, covered up or defaced in any way.
- Any goods or parts thereof, which may require repair or replacement, shall be repaired or replaced (at the discretion of Nickel Electro Ltd) at the works of Nickel Electro Ltd. The product to be repaired shall be delivered carriage paid back to Nickel Electro Ltd by the customer at the Customerās risk and expense. Any such goods or parts will be delivered by Nickel Electro Ltd to the Customer free within the United Kingdom but if required to be borne by the Customer. All faulty parts removed from the equipment will become Nickel Electro Ltdās property. Any other repairs or work by Nickel Electro Ltd will be carried out under the terms and conditions for specialist engineers currently in force.
- In the event of replacement with a new or reconditioned model, the replacement unit will continue the warranty period of the original equipment.
- If any goods or parts thereof are returned unnecessarily all cost involved, including a charge for inspection, handling and the return carriage must be paid by the sender. In no circumstances shall any of the goods be returned to Nickel Electro Ltd without its prior written consent.
- Please retain the original packaging over the warranty period.
Non Warranty Information
Spare parts shall be made available for a period of 3 years after a piece of equipment is discontinued.
Common Spare Parts
| Description | Part Number | Quantity | Where Used |
|---|---|---|---|
| Switch Blk/Grn 16A | ES0241 | 1 | NE1 Series |
| Power Entry Module | EX0854 | 1 | NE1 Series |
| Thermostat | SA01171 | 1 | NE1 Series |
| Fuse 2.5A | EF0095 | 2 | NE1-2.5 and NE1-4 |
| Fuse 5A | EF0096 | 2 | NE1-8 |
| Fuse 6.3A | EF0094 | 2 | NE1-14 |
| Fuse 7A | EF0436 | 1 | NE1-8 and NE1-14 |
| Fuse 8A | EF0632 | 2 | NE1-22 and NE1-28 |
| Fuse 10A | EF0335 | 1 | NE1-22 and NE1-28 |
| Circuit Breaker 15A | EX0550 | 2 | NE1-56 |
| 800W Matt Element | EX0859 | 1 | NE1-8, NE1-14 and NE1-56 baths |
| 1500W Matt Element | EX0826 | 1 | NE1-22 and NE1-28 |
| 250W Matt Element | EX0725 | 1 | NE1-2.5 |
| 400W Matt Element | EE0562 | 1 | NE1-4 |
Portable Appliance Testing
These tests should be conducted by a qualified person.
DO NOT PAT test the waterbath unless it contains water.
DO NOT Flash Test!!
Accessories For the NE1 Series Unstirred and Shallow Baths
GRP Lids
GL1-4
GL1-8
GL1-14
GL1-22
Hinged Gable Lid to suit 2 and 4 Litre Unstirred Baths
Hinged Gable Lid to suit 8 Litre Unstirred Baths
Hinged Gable Lid to suit 14 Litre Unstirred Baths
Hinged Gable Lid to suit 22 and 28 Litre Unstirred Baths
GL1-4AF
GL1-8AF
GL1-14AF
GL1-22AF
Lift Off Gable Lid to suit 2 and 4 Litre Unstirred Baths
Lift Off Gable Lid to suit 8 Litre Unstirred Baths
Lift Off Gable Lid to suit 14 Litre Unstirred Baths
Lift Off Gable Lid to suit 22 and 28 Litre Unstirred Baths
Stainless Steel Lids
SL1-4
SL1-8
SL1-14
SL1-14H
SL1-22
SL1-22H
SL1-56
SLR2-8
SLR1-14
SLR2-14
SLR1-22
LD-14
LD-22
Gabled Lid to suit 2 and 4 Litre Unstirred Baths
Gabled Lid to suit 8 Litre Unstirred Baths
Gabled Lid to suit 14 Litre Unstirred Baths
Hinged Gable Lid to suit 14 Litre Unstirred Baths
Gabled Lid to suit 22 and 28 Litre Unstirred Baths
Hinged Gable Lid to suit 22 and 28 Litre Unstirred Baths
Hinged Gable Lid to suit 56 Litre Unstirred Baths
2 x 114mm Ringed Lid to suit 7 Litre Unstirred Baths
4 x 114mm Ringed Lid to suit 14 and 18 Litre Unstirred Baths
6 x 83mm Ringed Lid to suit 14 and 18 Litre Unstirred Baths
6 x 105mm Ringed Lid to suit 28 Litre Unstirred Baths
Flat Lift Off Lid to suit 14 and 18 Litre Unstirred Baths
Flat Lift Off Lid to suit 28 Litre Unstirred Baths
Stainless Steel Test Tube Racks
Dimensions: 270 x 70 x 138mm
Quantity Req: 2/4 Litre = 1 Rack, 8 Litre = 2 Racks, 14 Litre = 4 Racks, 22 and 28 Litre = 6
Racks, 56 Litre = 12 Racks
6870
6871
6872
6873
6875
6900
26 Hole x 17mm Diameter
16 Hole x 26mm Diameter
36 Hole x 13mm Diameter
18 Hole x 19mm Diameter/suitable for 1.5ml Microtubes
50 Hole x 32mm Diameter/suitable for 50ml Falcon Tubes. Dimensions : 260 x 290 x 170mm
12 Hole x 32mm Diameter
Polypropylene Spheres
BP0368
Pack of 200
| DECLARATION OF CONFORMITY | |
| We herewith confirm the following product: NE1 Series Unstirred Thermostatic Baths. | |
| Conforms with the requirements outlined by the following European Directives: | Conforms with the requirements outlined in the following United Kingdom Directives: |
| Low Voltage Directive 2014/35/EU | Electromagnetic Compatibility Regulations 2016 |
| EMC Directive 2014/30/EU | Electrical Equipment (Safety) Regulations 2016 |
| RoHS Directive 2011/65/EU | RoHS Directive 2012/65/EU |
| Conforms with the requirements of the following standards: | |
| BS EN 61010-1: 2010 | Safety requirements for electrical equipment for measurement, control and laboratory use. |
| BS EN 61010-2-010: 2014 | |
| BS EN 61326-1: 2013 | Electrical equipment for measurement, control and laboratory use - EMC requirements. |
| Conforms with the requirements of the following standards: | |
| Nickel Electro Limited Oldmixon Crescent Weston super Mare North Somerset BS24 9BL United Kingdom t 01934 626691 f 01934 630300 e info@nickel-electro.co.uk |
FINAL INSPECTION AND ELECTRICAL SAFETY TEST REPORT.
BX0842 Issue 24: June 2025